M-3000A   Moving Die Rheometer

Moving Die Rheometer

  Vulcanization is the process that modifies the polymer by forming crosslink between individual polymer chains. This is one of the most important processes in manufacturing which improves the physical proprieties to the final rubber product.

  The rheometer can simultaneously determine the iscosity, elasticity, hysteresis loss and curing rate during the ulcanization process of the rubber compound. Important vulcanization characteristics such as torque, cure rate index and curing time can be measured.

產品規範: ASTM-D5289、GB/T 16584、ISO-6502

Model M-3000A M-3000FA
Die type Sealed biconical type
Oscillation frequency 100 cpm
Oscillation amplitude ±0.5°, ±1°, ±2° or ±3° (please specify)
Temperature control PC based software & PID
Temperature range Ambient temperature to 200℃
Temperature accuracy ±0.3℃
Temperature resolution 0.1℃
Grade of temp. sensor Class A
Heating up time 30℃ to 190℃<5 min 30 sec
175℃ to 190℃<1 min
Temperature recovery time Takes less than 50 sec to reach set value after placing specimen and closing die
Temperature overshoot Less than set value+0.4°C after closing die
Torque range Elasticity S`: 1~200 lb-in
Viscosity S": 0.3~200 lb-in
Torque unit kgf-cm、 lbf-in、N-m、dN-m
Torque resolution 0.001 dN-m
Pressure range N/A 0~1000 kg
Force unit N/A kgf、lbf、N、dN
Force resolution N/A 0.001 dN-m
Test time range 0~99 hours
Pneumatic pressure 3.6 kg/cm2 (0.36 MPa) prepared by customer
Dimension(W×D×H) 68×80×130cm
Weight (approx.) 220 kg
Power 1∮, AC220V±10%, 50/60Hz±3Hz, 10A

 


*Specification and content are subjected to change without notice.

Download the full PDF file

Download

Products